Miniaturized paper-based analytical device for the portable analysis of phyto-cannabinoids in plant and oral fluids

Anal Bioanal Chem. 2023 Nov 4. doi: 10.1007/s00216-023-05013-x. Online ahead of print.

ABSTRACT

In this work, a low-cost and eco-friendly paper-based analytical device (PAD) method is described for the determination of phyto-cannabinoids in cannabis and oral fluids based on a simple colorimetric reaction. The PAD was able to distinguish tetrahydrocannabinol (THC)- and cannabidiol (CBD)-rich plant samples by using 4-aminophenol (4-AP) and later on to quantify total phyto-cannabinoid content (THC + CBD + CBN) in plant and oral fluids by using the Fast Corinth V reagent. The chemical and physical properties regarding paper type and reagent concentration in the PAD were optimized to achieve the best analytical performance. After that, analytical features were obtained, including a linear range of 0.01-0.1 mg mL-1, a limit of detection (LOD) of 0.003 mg mL-1, and a suitable precision, expressed as relative standard deviation (RSD) lower than 10%. Furthermore, no significant interferences were observed in colorimetric reactions when tea, herbs, and drug samples were analyzed. Additionally, the PAD proved color stability up to 1 month after the sampling at 25 °C. The developed PAD was suitable for determining total phyto-cannabinoid content in plants and oral fluids, obtaining good results compared to GC-MS. Overall, this method showed good reliability resulting in an operational on-site device for drug monitoring.

PMID:37924377 | DOI:10.1007/s00216-023-05013-x